Page 2: General Information
General Information The spectrophotometer described in this manual is designed to be used by properly trained personnel in a suitable equipped laboratory. For the correct and safe use of this spectrophotometer it is essential for laboratory personnel to follow generally accepted safe procedures in addition to the safety precautions called for in this manual.
-
Page 3: Working Principle
Radio Interference For compliance with the EMC standards referred to in the EC Declaration of Conformity, ® it is necessary that only shielded cables supplied by UNICO are used when connecting the instrument to computers and accessories. Introduction ® UNICO 1200/1201 Spectrophotometers are single beam, general purpose instruments ®…
-
Page 4: Installation
Model 1200 Model 1201 Wavelength Range 325~1000 nm Spectral Bandpass 5 nm Wavelength Accuracy ± 2 nm Wavelength Repeatability ± 1 nm Stray Radiant Energy < 0.5%T at 340 and 400 nm Photometric Range 0~125%T 0~2.0 Abs 0~1999C (0~1999 Factor) Photometric Accuracy ±…
-
Page 5
Mode 0A/100%T Print/Enter Sample compartment Wavelength Knob Figure-3 UNICO ® 1200 Spectrophotometer MODE Button: There are four modes. T mode is transmittance mode; A mode is absorbance mode; C mode is to measure unknown sample concentration through a standard solution;… -
Page 6: Basic Operation
Turn on the spectrophotometer by turning on the Power Switch (IO). Allow 15 minutes for the instrument to warm up. ® Even through the 0%T is pre-set at UNICO ; it may shift due to unexpected factors like temperature change. To check 0%T, simply insert the Black Block (item number: S-1200-111, used to check the 0%T) into the Cuvette Holder and close the lid.
-
Page 7: Maintenance
follow step 2 above to set the concentration of the standard to the reduced value. 8. With the standard concentration set, determine the concentration values of samples with unknown concentration by inserting the sample cuvette into the Sample Compartment and reading the value direct from the Digital Display. 9.
-
Page 8
3. Loosen-up 2 lamp holding screws pointed by arrows on the image below. Remove damaged bulb and insert the new bulb all the way in to the holder. Lightly tighten the holding screws but not all the way. 4. Plug the power cord back and turn the instrument on. Set the wavelength to 340nm. Avoid to stair directly on to the bulb. -
Page 9
Wavelength Calibration Check Normally the UNICO ® 1200 Series spectrophotometer retains its wavelength calibration indefinitely. However if the instrument receives a severe shock or is abused, use the following methods to check wavelength calibration. Please note that this test requires ®… -
Page 10
It should be within 527~531 nm. Absorbance Accuracy Check Specification: ± 1% at 1A The absorbance accuracy should be checked against a set of neutral density filters ® accurately calibrated to the NIST standards. Contact your UNICO representative for more information (800-588-9776 or 732-274-1155). -
Page 11: Troubleshooting
Specification: Less than 0.5%T at 340 nm by ASTM E 387 The stray light should be checked against a set of filters accurately calibrated to the ® NIST standards. Contact your UNICO representative for more information (800-588- 9776 or 732-274-1155).
-
Page 12
Bubbles or particles in solution Check sample preparation and analytical procedure Instrument out of electronic calibration Call an authorized service engineer UNICO Authorized Service Contact Information UNICO 182 Ridge Road, Suite E, Dayton, NJ 08810, USA Tel.: 800-588-9776 or 732-274-1155 Fax: 732-274-1151 Web: www.unicosci.com…
This manual is also suitable for:
12001201
- Manuals
- Brands
- Unico Manuals
- Measuring Instruments
- 1200
- User manual
-
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Quick Links
1200 SERIES
SPECTROPHOTOMETER
USER’S MANUAL
United Products & Instruments, Inc.
182 Ridge Road, Suite E Dayton, NJ 08810, USA
TEL: 732-274-1155 FAX: 732-274-1151
www.unicosci.com sales@unicosci.com
1
Related Manuals for Unico 1200 Series
Summary of Contents for Unico 1200 Series
-
Page 1
1200 SERIES SPECTROPHOTOMETER USER’S MANUAL United Products & Instruments, Inc. 182 Ridge Road, Suite E Dayton, NJ 08810, USA TEL: 732-274-1155 FAX: 732-274-1151 www.unicosci.com sales@unicosci.com… -
Page 2: General Information
General Information The spectrophotometer described in this manual is designed to be used by properly trained personnel in a suitable equipped laboratory. For the correct and safe use of this spectrophotometer it is essential for laboratory personnel to follow generally accepted safe procedures in addition to the safety precautions called for in this manual.
-
Page 3: Working Principle
Radio Interference For compliance with the EMC standards referred to in the EC Declaration of Conformity, ® it is necessary that only shielded cables supplied by UNICO are used when connecting the instrument to computers and accessories. Introduction ® UNICO 1200/1201 Spectrophotometers are single beam, general purpose instruments ®…
-
Page 4: Installation
Model 1200 Model 1201 Wavelength Range 325~1000 nm Spectral Bandpass 5 nm Wavelength Accuracy ± 2 nm Wavelength Repeatability ± 1 nm Stray Radiant Energy < 0.5%T at 340 and 400 nm Photometric Range 0~125%T 0~2.0 Abs 0~1999C (0~1999 Factor) Photometric Accuracy ±…
-
Page 5
Mode 0A/100%T Print/Enter Sample compartment Wavelength Knob Figure-3 UNICO ® 1200 Spectrophotometer MODE Button: There are four modes. T mode is transmittance mode; A mode is absorbance mode; C mode is to measure unknown sample concentration through a standard solution;… -
Page 6: Basic Operation
Turn on the spectrophotometer by turning on the Power Switch (IO). Allow 15 minutes for the instrument to warm up. ® Even through the 0%T is pre-set at UNICO ; it may shift due to unexpected factors like temperature change. To check 0%T, simply insert the Black Block (item number: S-1200-111, used to check the 0%T) into the Cuvette Holder and close the lid.
-
Page 7: Maintenance
follow step 2 above to set the concentration of the standard to the reduced value. 8. With the standard concentration set, determine the concentration values of samples with unknown concentration by inserting the sample cuvette into the Sample Compartment and reading the value direct from the Digital Display. 9.
-
Page 8
3. Loosen-up 2 lamp holding screws pointed by arrows on the image below. Remove damaged bulb and insert the new bulb all the way in to the holder. Lightly tighten the holding screws but not all the way. 4. Plug the power cord back and turn the instrument on. Set the wavelength to 340nm. Avoid to stair directly on to the bulb. -
Page 9
Wavelength Calibration Check Normally the UNICO ® 1200 Series spectrophotometer retains its wavelength calibration indefinitely. However if the instrument receives a severe shock or is abused, use the following methods to check wavelength calibration. Please note that this test requires ®… -
Page 10
It should be within 527~531 nm. Absorbance Accuracy Check Specification: ± 1% at 1A The absorbance accuracy should be checked against a set of neutral density filters ® accurately calibrated to the NIST standards. Contact your UNICO representative for more information (800-588-9776 or 732-274-1155). -
Page 11: Troubleshooting
Specification: Less than 0.5%T at 340 nm by ASTM E 387 The stray light should be checked against a set of filters accurately calibrated to the ® NIST standards. Contact your UNICO representative for more information (800-588- 9776 or 732-274-1155).
-
Page 12
Bubbles or particles in solution Check sample preparation and analytical procedure Instrument out of electronic calibration Call an authorized service engineer UNICO Authorized Service Contact Information UNICO 182 Ridge Road, Suite E, Dayton, NJ 08810, USA Tel.: 800-588-9776 or 732-274-1155 Fax: 732-274-1151 Web: www.unicosci.com…
This manual is also suitable for:
12001201
The spectrophotometer described in this manual is designed to be used by properly
trained personnel in a suitable equipped laboratory. For the correct and safe use of this
spectrophotometer it is essential for laboratory personnel to follow generally accepted
safe procedures in addition to the safety precautions called for in this manual.
The inside of the power supply unit is a hazardous area and its cover should not be
removed under any circumstances. ANY Servicing must be done by an authorized
person.
Some of the chemicals you use in the spectrophotometer may be corrosive, flammable,
radioactive, toxic, and/or potentially infective. Care should be taken to follow the normal
laboratory procedures for handling chemicals and samples. Please carefully read the
Safety, Electrical, Warning, Performance and Radio Interference instructions below.
Safety
This spectrophotometer has been designed and tested in accordance with EN 61326-1:
1997 Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use standard (EMC Requirements). The spectrophotometer has been tested
and supplied in a safe condition.
Read the following before installing and using the instrument and its accessories.
Electrical
Before switching on the spectrophotometer, make sure it is set to the voltage of the
local power supply (see Installation section).
The main plug shall be inserted in a socket provided with a protective earth grounded
contact. The protective action must not be negated using an extension cord without a
protective conductor.
Warning
Any interruption of the protective conductor inside or outside the spectrophotometer or
disconnection of the protective earth terminal is likely to make the spectrophotometer
dangerous. Intentional interruption is prohibited. Whenever it is likely that the protection
has been impaired, the spectrophotometer shall be made inoperative and be secured
against any unintended operation.
The protection is likely to be impaired if, for example, the spectrophotometer
• Shows visible damage
• Fails to perform the intended measurements
• Has been subjected to prolonged storage under unfavorable conditions
• Has been subjected to severe transport stresses
Performance
Carry out performance checks with reference to wavelength and absorbance accuracy
to ensure that the spectrophotometer is working within its specification, especially when
making measurements of an important nature. Performance checks are detailed in this
manual.
2
З
АО НПО «КРИСМАС-Центр»
109316, РОССИЯ, Москва,
Остаповский пр., д. 13, стр.1, 102
т/ф. (095) 795-24-98
E-mail:christmasplus@mtu-net.ru
http://www.christmas-plus.ru
ФОТОМЕТР ЮНИКО 1200 (1201)
Новая модель фотометра особой надежности — ЮНИКО 1201, предназначена для измерения коэффициентов пропускания, оптической плотности и концентрации растворов. Однолучевой спектрофотометр специально разработан для широкого использования в лабораториях всех отраслей промышленности, решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
По техническим характеристикам и возможностям ЮНИКО 1200(1201) полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК и КФК-2,-3,-5 и др. Например, спектрофотометр обеспечивает определение содержания веществ в различных растворах – меди, железа, хлора, серебра; определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче; мочевины, общего белка, щелочей, фосфатов в химических растворах.
Фотометр ЮНИКО 1200 (1201) работает в спектральном интервале 325 – 1000 нм.
Прибор специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований.
Рабочая длина кювет 5-10-20-30-40-50 мм. Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта спектрофотометра КФК-3.
Фотометр обслуживается теми же сервисными службами, что и отечественные приборы, что обеспечивает минимальные сроки сервисного обслуживания. Прибор не требует специальных пуско-наладочных работ, допускает простую замену лампы без дополнительной юстировки.
Прибор готов к использованию непосредственно после доставки в лабораторию. Встроенный RS-232С порт для передачи данных в персональный компьютер позволяет работать со специализированным программным обеспечением.
Метрологическая поддержка.
На основании положительных результатов испытаний спектрофотометр ЮНИКО модель 1200(1201) внесен как тип в государственный реестр средств измерений РФ и зарегистрирован под № 24795, выдан сертификат Госстандарта РФ № 14767.
Спектрофотометры ЮНИКО 1200(1201) имеют описание типа средств измерений, методику поверки, руководство по эксплуатации на русском языке.
Приборы подлежат ежегодной поверке силами ЦСМ на местах с использованием стандартных поверочных средств (как КФК).
Технические характеристики |
Модель 1200 (1201) |
Спектральный диапазон длин волн | 325-1000нм |
Полоса пропускания | 5нм |
Погрешность установки длины волны, не более | 2 нм |
Повторяемость установки длины волны | 1нм |
Рассеянный свет (помехи лучистой энергии) | <0.5%Т при 340 и 400 нм |
Фотометрический диапазон: — коэффициент пропускания (Т)
— оптическая плотность (А) |
От 0 до 125 %
От 0 до 2.0 |
Диапазон значений концентрации | от 0 до 1999 С |
Погрешность определения коэффициента пропускания, не более | 1.0%Т |
Рабочая длина кювет | 5-10-20-30-40-50мм |
Напряжение питающей сети | 220 В 10 %, 50 Гц |
Мощность | 200 Вт |
Габариты (ширина * длина * высота); масса | 408*308*185 (мм); 7 кг |
Стандартный комплект поставки
Фотометр 1
Шнур питания 1
Чехол от пыли 1
Лампа галогеновая запасная 1
Заглушка — кюветодержатель
для кюветы 10 мм , контрольных
светофильтров и для компенсации темнового тока 1
Руководство по эксплуатации на русском языке 1
Сертификат 1
Паспорт 1
Методика поверки 1
Свидетельство о первичной поверке 1
Дополнительно (по заказу) поставляются: кюветы 5, 10, 20, 30, 40, 50 мм, программное обеспечение.
Рекомендации.
Десятки спектрофотометров ЮНИКО 1201 работают на предприятиях России (Водоканал г. Омск, Красноярск, Павловск Лен. Обл., ИркутскЭнерго, Калининская АЭС, Омский НПЗ, Ефремовский глюкозо-паточный завод, Орловский ГТУ, Новосибирский завод химконцентратов и др.).
Далее указаны некоторые предприятия, где установлены фотометры «Юнико 1200 (1201)»:
Водоканал г. Санкт-Петербург – 3 шт. , Сыктывкарский водоканал , Тольяттиазот, Бокситогорский глиноземный комбинат, Выксунский металлургический комбинат,
Орловский ГосТехУниверситет, Ефремовский Глюкозо-Паточный Завод (г. Ефремов, Тульская обл.), Калининская АЭС – 7 шт., ИркутскЭнерго – 3 шт., Тюменьэнерго, Казанский Государственный Энергетический Университет, Оренбургский Государственный Аграрный Университет, Сахалимморнефтегаз, Иркутский Государственный Университет, ООО «Мицар-Омск», Иткульский Спиртозавод и другие.
Пользователи отмечают высокую надежность прибора и удобство в работе.
Спектрофотометр используется для контроля состава воды, почвы, воздуха в экологических и санитарногигиенических исследованиях, а также при анализе сырья, готовой продукции, металлов и сплавов, химической продукции и т.д.
Однолучевой спектрофотометр специально разработан для широкого использования в лабораториях всех отраслей промышленности, решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
По техническим характеристикам и возможностям ЮНИКО 1200(1201) полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК и КФК-2,-3,-5 и др. Например, спектрофотометр обеспечивает определение содержания веществ в различных растворах – меди, железа, хлора, серебра; определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче; мочевины, общего белка, щелочей, фосфатов в химических растворах.
Фотометр ЮНИКО 1200 (1201) работает в спектральном интервале 325 – 1000 нм.
Прибор специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований.
Рабочая длина кювет 5-10-20-30-40-50 мм. Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта спектрофотометра КФК-3.
Фотометр обслуживается теми же сервисными службами, что и отечественные приборы, что обеспечивает минимальные сроки сервисного обслуживания. Прибор не требует специальных пуско-наладочных работ, допускает простую замену лампы без дополнительной юстировки.
Прибор готов к использованию непосредственно после доставки в лабораторию. Встроенный RS-232С порт для передачи данных в персональный компьютер позволяет работать со специализированным программным обеспечением.
Технические характеристики модель 1201
Спектральный диапазон длин волн | 325-1000нм |
Полоса пропускания | 5нм |
Погрешность установки длины волны, не более | 2 нм |
Повторяемость установки длины волны | 1нм |
Рассеянный свет (помехи лучистой энергии) | <0.5%Т при 340 и 400 нм |
Фотометрический диапазон: — коэффициент пропускания (Т) — оптическая плотность (А) |
От 0 до 125 % От 0 до 2.0 |
Диапазон значений концентрации | от 0 до 1999 С |
Погрешность определения коэффициента пропускания, не более | 1.0%Т |
Рабочая длина кювет | 5-10-20-30-40-50мм |
Напряжение питающей сети | 220 В ± 10 %, 50 Гц |
Мощность | 200 Вт |
Габариты (ширина * длина * высота); масса | 408*308*185 (мм); 7 кг |
Стандартный комплект поставки:
- Фотометр — 1 шт
- Шнур питания — 1 шт
- Чехол от пыли — 1 шт
- Лампа галогеновая запасная — 1 шт
- Заглушка — кюветодержатель для кюветы 10 мм , контрольных
- светофильтров и для компенсации темнового тока — 1шт
- Руководство по эксплуатации на русском языке — 1 шт
- Сертификат — 1 шт
- Паспорт — 1 шт
- Методика поверки — 1 шт
- Свидетельство о первичной поверке — 1 шт
- Программное обеспечение — 1 шт
- Дополнительно (по заказу) поставляются: кюветы 5, 10, 20, 30, 40, 50 мм.
Интересные материалы
Наши новости
Наши опросы
Артикул: 00000024849
Цена опт.: 152 830 руб.
В том числе НДС 20%
Описание
Спектрофотометр ЮНИКО 1200 (1201) предназначен для измерения коэффициентов пропускания, оптической плотности и концентрации растворов. Спектрофотометр Юнико 1200 используется в лабораториях всех отраслей промышленности, для решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
Спектрофотометр ЮНИКО 1201 специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований. По техническим характеристикам и возможностям спектрофотометр ЮНИКО 1201 полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК, КФК-3, КФК-5М и др. Например, спектрофотометр обеспечивает определение содержания веществ в различных растворах — меди, железа, хлора, серебра; определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче; мочевины, общего белка, щелочей, фосфатов в химических растворах. Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта фотометра КФК-3.
Компания НВ-Лаб предлагает специальные фотометрические наборы, в комплект входит описания фотометрических методик, оригинальные ГОСТЫ, набор специальных реагентов.
Документы и сертификаты
Технические характеристики
Спектральный |
315 — 1000 |
Диапазон измерений |
1 — 99 |
Диапазон показаний |
0 — 125 |
Диапазон измерений |
0,01 — 2,0 |
Диапазон показаний |
-0,1 — 2,5 |
Пределы допускаемой |
±1,0 |
Пределы допускаемой |
±2,0 |
Выделяемый |
5,0 |
Уровень рассеянного |
0,5 |
Дрейф показаний, |
— |
Габаритные размеры |
408*308* 185 |
Масса, кг |
6,5 |
Средний срок |
8 |
Потребляемая |
80 |
Напряжение питания |
220 (+15 -20)% |
Условия |
|
диапазон температур |
15 |
диапазон |
20 |
диапазон |
84 |
Транспортные характеристики
Онлайн-заявка
Если вы заинтересованы в приобретении товара «Спектрофотометр Юнико 1201», или хотите задать вопрос, то сделайте это прямо здесь:
— физическое лицо,
— юридическое лицо
Уважаемые покупатели, информируем Вас, что производитель может изменить цвет, внешний вид и характеристики товара без дополнительного уведомления продавца (НВ-Лаб). Поэтому размещенные на нашем сайте характеристики и фотографии являются справочными. Характеристики и внешний вид купленного товара иногда могут отличаться от опубликованных. Мы стараемся поддерживать описания в актуальном состоянии и обновляем информацию по мере получения её от производителей.
Вся размещённая на сайте информация опубликована исключительно для ознакомительных целей и ни при каких условиях не является публичной офертой.
The spectrophotometer described in this manual is designed to be used by properly
trained personnel in a suitable equipped laboratory. For the correct and safe use of this
spectrophotometer it is essential for laboratory personnel to follow generally accepted
safe procedures in addition to the safety precautions called for in this manual.
The inside of the power supply unit is a hazardous area and its cover should not be
removed under any circumstances. ANY Servicing must be done by an authorized
person.
Some of the chemicals you use in the spectrophotometer may be corrosive, flammable,
radioactive, toxic, and/or potentially infective. Care should be taken to follow the normal
laboratory procedures for handling chemicals and samples. Please carefully read the
Safety, Electrical, Warning, Performance and Radio Interference instructions below.
Safety
This spectrophotometer has been designed and tested in accordance with EN 61326-1:
1997 Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use standard (EMC Requirements). The spectrophotometer has been tested
and supplied in a safe condition.
Read the following before installing and using the instrument and its accessories.
Electrical
Before switching on the spectrophotometer, make sure it is set to the voltage of the
local power supply (see Installation section).
The main plug shall be inserted in a socket provided with a protective earth grounded
contact. The protective action must not be negated using an extension cord without a
protective conductor.
Warning
Any interruption of the protective conductor inside or outside the spectrophotometer or
disconnection of the protective earth terminal is likely to make the spectrophotometer
dangerous. Intentional interruption is prohibited. Whenever it is likely that the protection
has been impaired, the spectrophotometer shall be made inoperative and be secured
against any unintended operation.
The protection is likely to be impaired if, for example, the spectrophotometer
• Shows visible damage
• Fails to perform the intended measurements
• Has been subjected to prolonged storage under unfavorable conditions
• Has been subjected to severe transport stresses
Performance
Carry out performance checks with reference to wavelength and absorbance accuracy
to ensure that the spectrophotometer is working within its specification, especially when
making measurements of an important nature. Performance checks are detailed in this
manual.
2
- Manuals
- Brands
- Unico Manuals
- Measuring Instruments
- 1200
- User manual
-
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Quick Links
1200 SERIES
SPECTROPHOTOMETER
USER’S MANUAL
United Products & Instruments, Inc.
182 Ridge Road, Suite E Dayton, NJ 08810, USA
TEL: 732-274-1155 FAX: 732-274-1151
www.unicosci.com sales@unicosci.com
1
Related Manuals for Unico 1200 Series
Summary of Contents for Unico 1200 Series
-
Page 1
1200 SERIES SPECTROPHOTOMETER USER’S MANUAL United Products & Instruments, Inc. 182 Ridge Road, Suite E Dayton, NJ 08810, USA TEL: 732-274-1155 FAX: 732-274-1151 www.unicosci.com sales@unicosci.com… -
Page 2: General Information
General Information The spectrophotometer described in this manual is designed to be used by properly trained personnel in a suitable equipped laboratory. For the correct and safe use of this spectrophotometer it is essential for laboratory personnel to follow generally accepted safe procedures in addition to the safety precautions called for in this manual.
-
Page 3: Working Principle
Radio Interference For compliance with the EMC standards referred to in the EC Declaration of Conformity, ® it is necessary that only shielded cables supplied by UNICO are used when connecting the instrument to computers and accessories. Introduction ® UNICO 1200/1201 Spectrophotometers are single beam, general purpose instruments ®…
-
Page 4: Installation
Model 1200 Model 1201 Wavelength Range 325~1000 nm Spectral Bandpass 5 nm Wavelength Accuracy ± 2 nm Wavelength Repeatability ± 1 nm Stray Radiant Energy < 0.5%T at 340 and 400 nm Photometric Range 0~125%T 0~2.0 Abs 0~1999C (0~1999 Factor) Photometric Accuracy ±…
-
Page 5
Mode 0A/100%T Print/Enter Sample compartment Wavelength Knob Figure-3 UNICO ® 1200 Spectrophotometer MODE Button: There are four modes. T mode is transmittance mode; A mode is absorbance mode; C mode is to measure unknown sample concentration through a standard solution;… -
Page 6: Basic Operation
Turn on the spectrophotometer by turning on the Power Switch (IO). Allow 15 minutes for the instrument to warm up. ® Even through the 0%T is pre-set at UNICO ; it may shift due to unexpected factors like temperature change. To check 0%T, simply insert the Black Block (item number: S-1200-111, used to check the 0%T) into the Cuvette Holder and close the lid.
-
Page 7: Maintenance
follow step 2 above to set the concentration of the standard to the reduced value. 8. With the standard concentration set, determine the concentration values of samples with unknown concentration by inserting the sample cuvette into the Sample Compartment and reading the value direct from the Digital Display. 9.
-
Page 8
3. Loosen-up 2 lamp holding screws pointed by arrows on the image below. Remove damaged bulb and insert the new bulb all the way in to the holder. Lightly tighten the holding screws but not all the way. 4. Plug the power cord back and turn the instrument on. Set the wavelength to 340nm. Avoid to stair directly on to the bulb. -
Page 9
Wavelength Calibration Check Normally the UNICO ® 1200 Series spectrophotometer retains its wavelength calibration indefinitely. However if the instrument receives a severe shock or is abused, use the following methods to check wavelength calibration. Please note that this test requires ®… -
Page 10
It should be within 527~531 nm. Absorbance Accuracy Check Specification: ± 1% at 1A The absorbance accuracy should be checked against a set of neutral density filters ® accurately calibrated to the NIST standards. Contact your UNICO representative for more information (800-588-9776 or 732-274-1155). -
Page 11: Troubleshooting
Specification: Less than 0.5%T at 340 nm by ASTM E 387 The stray light should be checked against a set of filters accurately calibrated to the ® NIST standards. Contact your UNICO representative for more information (800-588- 9776 or 732-274-1155).
-
Page 12
Bubbles or particles in solution Check sample preparation and analytical procedure Instrument out of electronic calibration Call an authorized service engineer UNICO Authorized Service Contact Information UNICO 182 Ridge Road, Suite E, Dayton, NJ 08810, USA Tel.: 800-588-9776 or 732-274-1155 Fax: 732-274-1151 Web: www.unicosci.com…
This manual is also suitable for:
12001201
The spectrophotometer described in this manual is designed to be used by properly
trained personnel in a suitable equipped laboratory. For the correct and safe use of this
spectrophotometer it is essential for laboratory personnel to follow generally accepted
safe procedures in addition to the safety precautions called for in this manual.
The inside of the power supply unit is a hazardous area and its cover should not be
removed under any circumstances. ANY Servicing must be done by an authorized
person.
Some of the chemicals you use in the spectrophotometer may be corrosive, flammable,
radioactive, toxic, and/or potentially infective. Care should be taken to follow the normal
laboratory procedures for handling chemicals and samples. Please carefully read the
Safety, Electrical, Warning, Performance and Radio Interference instructions below.
Safety
This spectrophotometer has been designed and tested in accordance with EN 61326-1:
1997 Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use standard (EMC Requirements). The spectrophotometer has been tested
and supplied in a safe condition.
Read the following before installing and using the instrument and its accessories.
Electrical
Before switching on the spectrophotometer, make sure it is set to the voltage of the
local power supply (see Installation section).
The main plug shall be inserted in a socket provided with a protective earth grounded
contact. The protective action must not be negated using an extension cord without a
protective conductor.
Warning
Any interruption of the protective conductor inside or outside the spectrophotometer or
disconnection of the protective earth terminal is likely to make the spectrophotometer
dangerous. Intentional interruption is prohibited. Whenever it is likely that the protection
has been impaired, the spectrophotometer shall be made inoperative and be secured
against any unintended operation.
The protection is likely to be impaired if, for example, the spectrophotometer
• Shows visible damage
• Fails to perform the intended measurements
• Has been subjected to prolonged storage under unfavorable conditions
• Has been subjected to severe transport stresses
Performance
Carry out performance checks with reference to wavelength and absorbance accuracy
to ensure that the spectrophotometer is working within its specification, especially when
making measurements of an important nature. Performance checks are detailed in this
manual.
2
З
АО НПО «КРИСМАС-Центр»
109316, РОССИЯ, Москва,
Остаповский пр., д. 13, стр.1, 102
т/ф. (095) 795-24-98
E-mail:christmasplus@mtu-net.ru
http://www.christmas-plus.ru
ФОТОМЕТР ЮНИКО 1200 (1201)
Новая модель фотометра особой надежности — ЮНИКО 1201, предназначена для измерения коэффициентов пропускания, оптической плотности и концентрации растворов. Однолучевой спектрофотометр специально разработан для широкого использования в лабораториях всех отраслей промышленности, решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
По техническим характеристикам и возможностям ЮНИКО 1200(1201) полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК и КФК-2,-3,-5 и др. Например, спектрофотометр обеспечивает определение содержания веществ в различных растворах – меди, железа, хлора, серебра; определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче; мочевины, общего белка, щелочей, фосфатов в химических растворах.
Фотометр ЮНИКО 1200 (1201) работает в спектральном интервале 325 – 1000 нм.
Прибор специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований.
Рабочая длина кювет 5-10-20-30-40-50 мм. Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта спектрофотометра КФК-3.
Фотометр обслуживается теми же сервисными службами, что и отечественные приборы, что обеспечивает минимальные сроки сервисного обслуживания. Прибор не требует специальных пуско-наладочных работ, допускает простую замену лампы без дополнительной юстировки.
Прибор готов к использованию непосредственно после доставки в лабораторию. Встроенный RS-232С порт для передачи данных в персональный компьютер позволяет работать со специализированным программным обеспечением.
Метрологическая поддержка.
На основании положительных результатов испытаний спектрофотометр ЮНИКО модель 1200(1201) внесен как тип в государственный реестр средств измерений РФ и зарегистрирован под № 24795, выдан сертификат Госстандарта РФ № 14767.
Спектрофотометры ЮНИКО 1200(1201) имеют описание типа средств измерений, методику поверки, руководство по эксплуатации на русском языке.
Приборы подлежат ежегодной поверке силами ЦСМ на местах с использованием стандартных поверочных средств (как КФК).
Технические характеристики |
Модель 1200 (1201) |
Спектральный диапазон длин волн | 325-1000нм |
Полоса пропускания | 5нм |
Погрешность установки длины волны, не более | 2 нм |
Повторяемость установки длины волны | 1нм |
Рассеянный свет (помехи лучистой энергии) | <0.5%Т при 340 и 400 нм |
Фотометрический диапазон: — коэффициент пропускания (Т)
— оптическая плотность (А) |
От 0 до 125 %
От 0 до 2.0 |
Диапазон значений концентрации | от 0 до 1999 С |
Погрешность определения коэффициента пропускания, не более | 1.0%Т |
Рабочая длина кювет | 5-10-20-30-40-50мм |
Напряжение питающей сети | 220 В 10 %, 50 Гц |
Мощность | 200 Вт |
Габариты (ширина * длина * высота); масса | 408*308*185 (мм); 7 кг |
Стандартный комплект поставки
Фотометр 1
Шнур питания 1
Чехол от пыли 1
Лампа галогеновая запасная 1
Заглушка — кюветодержатель
для кюветы 10 мм , контрольных
светофильтров и для компенсации темнового тока 1
Руководство по эксплуатации на русском языке 1
Сертификат 1
Паспорт 1
Методика поверки 1
Свидетельство о первичной поверке 1
Дополнительно (по заказу) поставляются: кюветы 5, 10, 20, 30, 40, 50 мм, программное обеспечение.
Рекомендации.
Десятки спектрофотометров ЮНИКО 1201 работают на предприятиях России (Водоканал г. Омск, Красноярск, Павловск Лен. Обл., ИркутскЭнерго, Калининская АЭС, Омский НПЗ, Ефремовский глюкозо-паточный завод, Орловский ГТУ, Новосибирский завод химконцентратов и др.).
Далее указаны некоторые предприятия, где установлены фотометры «Юнико 1200 (1201)»:
Водоканал г. Санкт-Петербург – 3 шт. , Сыктывкарский водоканал , Тольяттиазот, Бокситогорский глиноземный комбинат, Выксунский металлургический комбинат,
Орловский ГосТехУниверситет, Ефремовский Глюкозо-Паточный Завод (г. Ефремов, Тульская обл.), Калининская АЭС – 7 шт., ИркутскЭнерго – 3 шт., Тюменьэнерго, Казанский Государственный Энергетический Университет, Оренбургский Государственный Аграрный Университет, Сахалимморнефтегаз, Иркутский Государственный Университет, ООО «Мицар-Омск», Иткульский Спиртозавод и другие.
Пользователи отмечают высокую надежность прибора и удобство в работе.
Спектрофотометр используется для контроля состава воды, почвы, воздуха в экологических и санитарногигиенических исследованиях, а также при анализе сырья, готовой продукции, металлов и сплавов, химической продукции и т.д.
Однолучевой спектрофотометр специально разработан для широкого использования в лабораториях всех отраслей промышленности, решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
По техническим характеристикам и возможностям ЮНИКО 1200(1201) полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК и КФК-2,-3,-5 и др. Например, спектрофотометр обеспечивает определение содержания веществ в различных растворах – меди, железа, хлора, серебра; определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче; мочевины, общего белка, щелочей, фосфатов в химических растворах.
Фотометр ЮНИКО 1200 (1201) работает в спектральном интервале 325 – 1000 нм.
Прибор специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований.
Рабочая длина кювет 5-10-20-30-40-50 мм. Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта спектрофотометра КФК-3.
Фотометр обслуживается теми же сервисными службами, что и отечественные приборы, что обеспечивает минимальные сроки сервисного обслуживания. Прибор не требует специальных пуско-наладочных работ, допускает простую замену лампы без дополнительной юстировки.
Прибор готов к использованию непосредственно после доставки в лабораторию. Встроенный RS-232С порт для передачи данных в персональный компьютер позволяет работать со специализированным программным обеспечением.
Технические характеристики модель 1201
Спектральный диапазон длин волн | 325-1000нм |
Полоса пропускания | 5нм |
Погрешность установки длины волны, не более | 2 нм |
Повторяемость установки длины волны | 1нм |
Рассеянный свет (помехи лучистой энергии) | <0.5%Т при 340 и 400 нм |
Фотометрический диапазон: — коэффициент пропускания (Т) — оптическая плотность (А) |
От 0 до 125 % От 0 до 2.0 |
Диапазон значений концентрации | от 0 до 1999 С |
Погрешность определения коэффициента пропускания, не более | 1.0%Т |
Рабочая длина кювет | 5-10-20-30-40-50мм |
Напряжение питающей сети | 220 В ± 10 %, 50 Гц |
Мощность | 200 Вт |
Габариты (ширина * длина * высота); масса | 408*308*185 (мм); 7 кг |
Стандартный комплект поставки:
- Фотометр — 1 шт
- Шнур питания — 1 шт
- Чехол от пыли — 1 шт
- Лампа галогеновая запасная — 1 шт
- Заглушка — кюветодержатель для кюветы 10 мм , контрольных
- светофильтров и для компенсации темнового тока — 1шт
- Руководство по эксплуатации на русском языке — 1 шт
- Сертификат — 1 шт
- Паспорт — 1 шт
- Методика поверки — 1 шт
- Свидетельство о первичной поверке — 1 шт
- Программное обеспечение — 1 шт
- Дополнительно (по заказу) поставляются: кюветы 5, 10, 20, 30, 40, 50 мм.
- Измерение коэффициентов пропускания, оптической плотности и концентрации растворов.
Габаритные размеры (Д×Ш×В)
Вес
Питание
Спектральный диапазон
Потребляемая мощность
Диапазон оптической плотности
Коэффициент пропускания (Т)
Спектрофотометр ЮНИКО 1200
408×308×185 мм
7 кг
220/50 В/Гц
325…1000 нм
2 кВт
0,00…2,00
0…125%
Описание
Однолучевой спектрофотометр ЮНИКО 1200 создан для работы в спектральном интервале 325…1000 нм и предназначен для проведения измерений коэффициентов пропускания, а также оптической плотности и концентрации растворов.
Нефтехимия
анализ бензинов
Комплектация
- Спектрофотометр ЮНИКО 1200.
Мы поможем вам подобрать и купить необходимое оборудование для лаборатории. Компания
«МИЛЛАБ» работает с производителями напрямую, потому у нас объективные цены на всю
продукцию и устройства!
Вы просматривали
Аналитика
Аналитическое и научное оборудование
Общелабораторное оборудование
Лабораторная мебель
Биотех
Оборудование для микробиологии и фармацевтики
Расходные материалы
Питательные среды
Синтез
Оборудование для лабораторного, пилотного и крупнотоннажного синтеза
Технология
Испытание материалов
Термообработка
Вакуумное оборудование
Инжиниринг
Проектные решения для производства и лабораторий.
Предпроект, 3D- визуализация.
Сервис
Сервисная поддержка
Аттестация, валидация оборудования
Поставка запасных частей
_______________________________
Доброго времени суток, коллеги!
Убедительная просьба, если у кого нибудь есть в наличие МП, такие как
—Спектрофотометры UNICO, модели 1200, 1201 производства «United Products & Instruments, Inc», США. Методика поверки (БШ 2.853.021 МП)
—Спектрофотометры UVmini-1240, UV-2401PC, UV-1700, UV-1650PC, UV-2501PC, UV-3600, SolidSpec-3700, производства фирмы «SHIMADZU», Япония. Методика поверки
-Анализаторы общего азота и общего углерода элементные типа ТОПАЗ-NC (ЛШЮГ 413411.019 ДЛ)
В скобках я указал предположительные(наверняка не знаю — источник недостоверный) обозначения этих методик.
У нас есть средства поверки, а методик к сожалению нету.
Буду очень признателен за помощь!
Артикул: PE_282
Артикул: PE_282
Номер товара: 23616
Характеристики
Все характеристики
Спектрофотометр UNICO 1200
Минимальная сумма заказа в нашем магазине от 5000 руб.
Цены на нашем сайте актуальны при заказе от 20 000 рублей.
Мы предлагаем индивидуальные условия оплаты.
Заказывайте также по телефону или по электронной почте:
-
Фотометр Unico 1200 cнят с производства и заменен в 2008 году на ближайший аналог: фотометр Unico 1201
Спектрофотометры ЮНИКО 1200 и 1201 предназначены для измерения коэффициентов пропускания, оптической плотности и концентрации растворов.
Спектрофотометр Юнико 1200 используется в лабораториях всех отраслей промышленности, для решения аналитических задач в таких отраслях, как клиническая химия, биохимия, защита окружающей среды, энергетика, металлургические, химические, пищевые лаборатории, лаборатории воды и сточной воды, нефтехимии и в других сферах контроля качества.
Спектрофотометр ЮНИКО 1200 специально адаптирован для отечественных условий эксплуатации и выпускается с учетом российских лабораторных требований. По техническим характеристикам и возможностям ЮНИКО 1200 полностью заменяет широко применяемые в лабораторной практике фотоколориметры и фотометры фотоэлектрические типа ФЭК, КФК-3, КФК-5М и др.
Этот прибор обеспечивает определение содержания веществ в различных растворах:
- меди,
- железа,
- хлора,
- серебра;
- определение содержания сахара, билирубина, глюкозы, холестерина, креатина в крови и моче;
- мочевины, общего белка, щелочей, фосфатов в химических растворах.
Кюветное отделение аппарата переработано под российский стандарт. Предусмотрено использование кювет из комплекта фотометра КФК-3.
На основании положительных результатов испытаний спектрофотометр ЮНИКО 1200 внесен как тип в государственный реестр средств измерений. Также выдан сертификат Госстандарта РФ 14767.
Спектрофотометры ЮНИКО 1200 имеют:
- описание типа средств измерений, методику поверки;
- руководство по эксплуатации на русском языке;
- ЮНИКО подлежат ежегодной поверке силами ЦСМ на местах с использованием стандартных поверочных средств (как КФК).
Спектральный диапазон длин волн 325-1000 нм Полоса пропускания 5 нм Погрешность установки длины волны, не более 2 нм Повторяемость установки длины волны 1 нм Рассеянный свет (помехи лучистой энергии) 0.5%Т при 340 и 400 нм Диапазон значений концентрации от 0 до 1999 С Погрешность определения коэффициента пропускания, не более 1.0%Т Рабочая длина кювет 5-10-20-30-40-50 мм Напряжение питающей сети 220 В + 10 %, 50 Гц Габариты (ширина*длина*высота); масса 408*308*185 (мм); 7 кг
Упакуем ваш заказ
Мы надежно упакуем ваш заказ
Доставка
Быстрая и бережная доставка по всей России
Гарантия и качество
Официальная гарантия от производителя
Удобная оплата
Предлагаем индивидуальные условия оплаты