Vortex Flow Measuring System
Reliable flow measurement of gas, steam and liquids
Application
For the universal measurement of the volume flow
of gases, steam and liquids.
With integrated T-measurement, the mass flow of steam,
water (as per IAPWS-IF97 ASME), natural gas (as per
AGA NX-19), compressed air and other gases and liquids
can be measured.
Maximum range of applications thanks to:
• Fluid temperature range from –200…+400 °C
• Pressure ratings up to PN250/Cl 1500
Approvals for hazardous areas:
• ATEX, FM, CSA, TIIS
Connection to all common process control systems:
• HART, PROFIBUS PA, FOUNDATION Fieldbus
Relevant safety aspects:
• Pressure Equipment Directive
• Up to SIL 2
TI070D/06/en/12.05
71005481
Your benefits
The robust Prowirl sensor, tried and tested in over
100,000 applications offers:
• High resistance to:
– Vibrations (over 1 g in all axes)
– Temperature shocks (> 150 K/s)
– Contaminated media
– Water hammer
• No maintenance, no moving parts, no zero-point drift.
• Software initial settings save time and costs
In addition, Prowirl devices offer the following
possibilities:
• Complete saturated steam or liquid-mass measuring
point in one single device
• Calculation of the mass flow from the measured
variables volume flow and temperature in the
integrated flow computer
• External pressure value read-in for superheated steam
and gas applications (optional)
• External temperature value read-in for delta heat
measurement (optional)
Summary of Contents for Endress+Hauser Proline Prowirl 72F
(Ocr-Read Summary of Contents of some pages of the Endress+Hauser Proline Prowirl 72F Document (Main Content), UPD: 16 February 2023)
-
9, Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 9 Signal on alarm • Current output: error response can be selected (e.g. in accordance with NAMUR Recommendation NE 43) • Pulse output: error response can be selected • Status output: «not conducting» in event of fault (open circuit) Load A0001921 The area shaded gray refers to the permitted load (for HART: min. 250 Ω) The load is calculated as follows: R B…
-
8, Proline Prowirl 72F, 72W, 73F, 73W 8 Endress+Hauser Prowirl 73 Current output: • 4…20 mA with HART, • Full scale value and time constant (0…100 s) can be set Frequency output, pulse/status output: • Frequency output (optional): open collector, passive, galvanically isolated – Non-Ex, Ex d version: U max = 36 V, with 15 mA current limiting, R i = 500 Ω – Ex i and Ex n version: U max = 30 V, with 15 mA current limiting, R i = 500 Ω The puls…
-
16, Proline Prowirl 72F, 72W, 73F, 73W 16 Endress+Hauser Perforated plate flow conditioner A specially designed perforated plate flow conditioner, available from Endress+Hauser, can be installed if it is not possible to observe the inlet runs required. The flow conditioner is fitted between two piping flanges and centered with the mounting bolts. Generally, this reduces the inlet run required to 10 …
-
7, Endress+Hauser Proline Prowirl 72F Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 7 Output signal Prowirl 72 Current output: • 4…20 mA with HART, • Full scale value and time constant (0…100 s) can be set Pulse/status output: • Open collector, passive, galvanically isolated – Non-Ex, Ex d version: U max = 36 V, with 15 mA current limiting, R i = 500 Ω – Ex i and Ex n version: U max = 30 V, with 15 mA current limiting, R i = 50…
-
26, Proline Prowirl 72F, 72W, 73F, 73W 26 Endress+Hauser Dimensions of Prowirl 72F, Dualsens version Flanged version to: • EN 1092-1 (DIN 2501), Ra = 6.3…12.5 mm • Raised face to: – EN 1092-1 Form B1 (DIN 2526 Form C), PN 10…40, Ra = 6.3…12.5 mm, optional with groove to EN1091-1 Form D (DIN 2512 Form N) – EN 1092-1 Form B2 (DIN 2526 Form E), PN 63…100, Ra = 1.6…3.2 mm 1) 2) – DIN 2526 Form E, PN 1…
-
33, Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 33 Accessories • Spare parts as per separate price list • Wafer version mounting set • Pipe and wall mounting kit for transmitter remote version DK5WM-B • Replacement transmitter • Universal flow and energy computer RMC 621 • Steam computer RMS 621 • Flow conditioner • HART Communicator DXR 375 handheld terminal • Active barrier preline RN 221 N • Resist…
-
31, Endress+Hauser Proline Prowirl 72F Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 31 Certificates and approvals CE mark The device is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing the CE mark. C-tick The measuring system is in conformity with the EMC requirements of the Australian Communications Authority (ACA). Ex-approval • Ex i and Ex n: –ATEX/CENELEC II1/2G, EEx ia IIC T1…T6 (T1…T4 for P…
-
2, Proline Prowirl 72F, 72W, 73F, 73W 2 Endress+Hauser Table of contents Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Measured variable . . . …
-
1, TI070D/06/en/12.05 71005481 Technical Information Proline Prowirl 72F, 72W, 73F, 73W Vortex Flow Measuring System Reliable flow measurement of gas, steam and liquids Application For the universal measurement of the volume flow of gases, steam and liquids. With integrated T-measurement, the mass flow of steam, water (as per IAPWS-IF97 ASME), natural gas (as per AGA NX-19), compressed air and other gases and liquids can be measured. Maximum range of applications thanks to: • Fluid temp…
-
27, Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 27 Prowirl 72F Dualsens version to EN 1092-1 (DIN 2501) DN Pressure rating d D H L X Weight DIN/JIS [mm] [mm] [mm] [mm] [mm] [kg] 40 PN 40 43.1 150.0 303 200 31 16 PN 100 42.5 170.0 18 PN 160 41.1 170.0 50 PN 40 54.5 165.0 310 200 33 18 PN 63 54.5 180.0 20 PN 100 53.9 195.0 22 PN 160 52.3 195.0 80 PN 40 82.5 200.0 323 200 39 25 PN 63 81.7 215.0 27 PN 100 80.9 230.0 30 PN 160 76.3 230.0 100 PN 16 107.1 220.0 335 250 49 42PN…
-
24, Proline Prowirl 72F, 72W, 73F, 73W 24 Endress+Hauser Prowirl 72F, 73F to ANSI B16.5 DN Pressure rating d [mm] D [mm] H [mm] L [mm] X [mm] Weight [kg] ½» Schedule 40 Cl. 150 15.7 88.9 248 200 11.2 5 Cl. 300 15.7 95.0 14.2 Schedule 80 Cl. 150 13.9 88.9 11.2 Cl. 300 13.9 95.0 14.2 Cl. 600 2) 13.9 95.3 288 200 23 6 Cl. 1500 1) 14.0 120.6 310 262 22.3 13 Butt-weld 1) 14.0 21.3 310 248 — 9 1″ Schedule 40 Cl. 150 26.7 107.9 255 200 15.7 7 C…
-
4, Proline Prowirl 72F, 72W, 73F, 73W 4 Endress+Hauser The capacitive sensor The sensor of a vortex flowmeter has a major influence on the ability, robustness and reliability of the whole measuring system. The robust DSC sensor — with an integrated temperature sensor (Pt 1000) with Prowirl 73 — is burst-tested and vibration and temperature-shock-tested (temperature shocks of 150 K/s). The Prowirl uses the tried-and…
-
36, Proline Prowirl 72F, 72W, 73F, 73W International Head Quarter Endress+Hauser GmbH+Co. KG Instruments International Colmarer Str. 6 79576 Weil am Rhein Deutschland Tel. +49 76 21 9 75 02 Fax +49 76 21 9 75 34 5 www.endress.com [email protected] TI070D/06/en/12.05 71005481 FM+SGML 6.0 ProMoDo
… -
3, Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 3 Function and system design Measuring principle Vortex meters work on the principle of the Karman vortex street. When luid flows past a bluff body, vortices are alternately formed on both sides with opposite directions of rotation. These vortices each generate a local low pressure. The pressure fluctuations are recorded by the sensor and converted to…
-
14, Endress+Hauser Proline Prowirl 72F Proline Prowirl 72F, 72W, 73F, 73W 14 Endress+Hauser Piping insulation When insulating, please ensure that a sufficiently large area of the housing support is exposed. The uncovered part serves as a radiator and protects the electronics from overheating (or undercooling). The maximum insulation height permitted is illustrated in the diagrams. These apply equally to both the compact version and the sensor in the remote version. A0001868 1 = Flanged version 2 = Wafer versi…
-
13, Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 13 Operating conditions: installation Installation instructions Vortex meters require a fully developed flow profile as a prerequisite for correct volume flow measurement. For this reason, please note the following points when installing the device: Orientation The device can generally be installed in any position in the piping. In the case of liquids, there should be upward flow in vertical pipes…
Table of Contents for Endress+Hauser Proline Prowirl 72F:
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 33 Accessories • Spare parts as per separate price list • Wafer version mounting set • Pipe and wall mounting kit for transmitter remote version DK5WM-B • Replacement transmitter • Universal flow and energy computer RMC 621 • Steam computer RMS 621 • Flow conditioner • HART Communicator DXR 375 handheld terminal • Active barrier preline RN 221 N • Resistance thermometer Omnigrad TR10 (HART-enabled and burst-enabled
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 23 1) In contrast to the other versions, devices have a sensor in the bluff body. Only available for 72F. 2) Pressure ratings are in development for Prowirl 73. Prowirl 72F, 73F to EN 1092-1 (DIN 2501) DN Pressure rating d [mm] D [mm] H [mm] L [mm] X [mm] Weight [kg] 15 PN 40 17.3 95.0 248 200 16 5 PN 160 2) 17.3 105.0 288 200 23 7 PN 250 1) 14.0 130.0 310 248 26 15 Butt-weld 1) 14.0 21.3 310 248 — 9 25 PN 40 28.5 1
-
TI070D/06/en/12.05 71005481 Technical Information Proline Prowirl 72F, 72W, 73F, 73W Vortex Flow Measuring System Reliable flow measurement of gas, steam and liquids Application For the universal measurement of the volume flow of gases, steam and liquids. With integrated T-measurement, the mass flow of steam, water (as per IAPWS-IF97 ASME), natural gas (as per AGA NX-19), compressed air and other gases and liquids can be measured. Maximum range
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 27 Prowirl 72F Dualsens version to EN 1092-1 (DIN 2501) DN Pressure rating d D H L X Weight DIN/JIS [mm] [mm] [mm] [mm] [mm] [kg] 40 PN 40 43.1 150.0 303 200 31 16 PN 100 42.5 170.0 18 PN 160 41.1 170.0 50 PN 40 54.5 165.0 310 200 33 18 PN 63 54.5 180.0 20 PN 100 53.9 195.0 22 PN 160 52.3 195.0 80 PN 40 82.5 200.0 323 200 39 25 PN 63 81.7 215.0 27 PN 100 80.9 230.0 30 PN 160 76.3 230.0 100 PN 16 107.1 220.0 335 250 49 42PN 40 107.1 235.0
-
Proline Prowirl 72F, 72W, 73F, 73W 18 Endress+Hauser Fluid pressure Prowirl 72 Pressure-temperature curve to EN (DIN), stainless steel PN 10…40 → Prowirl 72W and 72F PN 63…250 → Prowirl 72F A0003238 Pressure-temperature curve to ANSI B16.5, stainless steel ANSI B16.5: Class 150…300 → Prowirl 72W and 72F Class 600…1500 → Prowirl 72F A0003402 -200 -100 0 20 40 60 80 100 120 140 160 0 100 200 300 400 °C bar PN 63…250 -200 -100 0 5 10 15 20 25 30 35 40 45 0 100 200 300 400 °C bar PN 10…40 PN 100 PN 1
-
Proline Prowirl 72F, 72W, 73F, 73W 16 Endress+Hauser Perforated plate flow conditioner A specially designed perforated plate flow conditioner, available from Endress+Hauser, can be installed if it is not possible to observe the inlet runs required. The flow conditioner is fitted between two piping flanges and centered with the mounting bolts. Generally, this reduces the inlet run required to 10 x DN with complete accuracy. A0001887 The pressure loss for flow conditioners is calc
-
Proline Prowirl 72F, 72W, 73F, 73W 2 Endress+Hauser Table of contents Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Measuring range . . . . . . . . . .
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 19 Pressure-temperature curve to JIS B2238 and butt-weld version, stainless steel: A0003404 Pressure-temperature curve to EN (DIN), ANSI B16.5 and JIS B2238, Alloy C-22 PN 16…40, Class 150…300, 10…20K → Prowirl 72F A0003395 Prowirl 73 Pressure-temperature curve to EN (DIN), stainless steel PN 10…40 → Prowirl 73W and 73F PN 63…160 → Prowirl 73F (in development) A0001922 • JIS B2238: 10…20K → Prowirl 72W an
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 25 1) In contrast to the other versions, devices have a sensor in the bluff body. Only available for 72F. 2) Pressure ratings are in development for Prowirl 73. 1) Pressure rating 40 K for Prowirl 73 in development 10″ Schedule 40 Cl. 150 254.5 406.4 375 380 48 92 Cl. 300 254.5 444.5 109 12″ Schedule 40 Cl. 150 304.8 482.6 398 450 60 143 Cl. 300 304.8 520.7 162 Prowirl 72F, 73F to JIS B2238 DN Pressure rating d
-
Proline Prowirl 72F, 72W, 73F, 73W 32 Endress+Hauser Other standards and guidelines • EN 60529: Degrees of protection by housing (IP code). • EN 61010: Protection measures for electrical equipment for measurement, control, regulation and laboratory procedures. • EN 61326/A1: Electromagnetic compatibility (EMC requirements). • NAMUR NE 21: Electromagnetic compatibility (EMC) o
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 17 Operating conditions: environment Ambient temperature range • Compact version: –40…+70 °C (EEx-d version: –40…+60 °C; ATEX II 1/2 GD version/dust ignition-proof: –20…+55 °C) Display can be read between –20 °C…+70 °C • Remote version: Sensor –40…+85 °C (ATEX II 1/2 GD version/dust ignition-proof: –20…+55 °C) Transmitter –40…+80 °
-
Proline Prowirl 72F, 72W, 73F, 73W 34 Endress+Hauser
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 11 Wiring (remote version) Note! • The remote version must be grounded. In doing so, the sensor and transmitter must be connected to the same potential matching system. A0001893 Connecting the remote version a = Connection compartment cover (transmitter) b = Connection compartment cover (sensor) c = Connecting cable (signal cable) d = Identical potential matching for sensor and transmitter e = Connect shieldi
-
Proline Prowirl 72F, 72W, 73F, 73W Endress+Hauser 31 Certificates and approvals CE mark The device is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing the CE mark. C-tick The measuring system is in conformity with the EMC requirements of the Australian Communications Authority (ACA). Ex-approval • Ex i and Ex n: –ATEX/CENELEC II1/2G, EEx ia IIC T1…T6 (T1…T4 for PROFIBUS PA and FOUNDATION Fieldbus) II1/2GD, EEx ia IIC T1…T6 (T1…T4 for PROFIBUS PA